SOFTWARE CONFIGURATION
25
4
4.3 SCANNER CONFIGURATION AND CALIBRATION
4.3.1 Scanner Parameter Configuration
From Genius™:
1. Double click the icon of one of the scanners to connect to it and modify its configuration.
Typically it is necessary to adjust only Code Definition (expected labels symbology and
length), allowing the scanner to read specific codes.
2. Copy the configuration from a device by pointing and right-clicking its icon.
3. Tick all the scanner checkboxes and Paste the configuration to all the devices by pointing
and right-clicking the icon of the first device.
4. Tick all the scanner checkboxes and press the Send icon on the GENIUS
TM
toolbar to
distribute the configuration to all the nodes.
5. Only for the scanners reading from FRONT and BACK
sides set the following: Reading Parameters > Reading
Condition > Skew 45°. This setting is needed every time
the labels are read with a skew angle of 45 degrees and
provides increased gain for the signal.
CAUTION
Copy and paste must only be used the first time, otherwise the parameter
differences will be overwritten effecting the working of the scanner.
4.3.2 Reading Performance Check
For each scanner to check:
1. Double click the icon of the scanner to check. Enter the SPY add-on
2. Press the button Test.
3. Place a label under the beam and make sure that it is read in the whole reading area with
a percentage > 95%.